Simulation for level Enter the level value to be simulated. The appropriate current is given by the
electronic insert.
Step Entry Cursor in Display at Significance
1 3
SERVICE/SIMULATION
Service/Simulation
26
SIM. LEVEL
Simulation for level
e.g. 77.06
Actual measured value is displayed
3 e.g. 35.00
35.00
Level to be simulated is entered
4 F4
[ENTER]
Confirms entry, value to be transmitted
Simulation for volume Enter the volume value to be simulated. The appropriate current is given by the
electronic insert.
Step Entry Cursor in Display to Significance
1 3
SERVICE/SIMULATION
Service/Simulation
27
SIM. VOLUME
Simulation of volume
e.g. 77.06
Actual measured value is displayed
3 e.g. 5.00
5.00
Volume to be simulated is entered
4 F4
[ENTER]
Confirms entry, value to be transmitted
Simulation for current Enter directly the current value to be simulated.
Step Entry Cursor in Display at Significance
1 3
SERVICE/SIMULATION
Service/Simulation
27
SIM. CURRENT
Simulation of current output
e.g. 17.02
Actual measured value is displayed
3 e.g. 8.00
8.00
Current to be simulated is entered
4 F4
[ENTER]
Confirms entry, value to be transmitted
Note!
The values entered during simulation are stored until simulation is again switched off.
Note!
6 Diagnosis and Troubleshooting Electronic insert FEC 12 (HART)