35-3000RK-LEL Combustilbe Gas Sample-Draw Detector • 17
NOTE: Make sure the bottom ferrule is laying at in the nut.
6. Insert the cone-shaped front ferrule on top of the bottom ferrule. Insert the ferrule so
the smaller end of the cone is facing up.
7. Screw the nut onto the tting by hand just until it stops turning.
8. Insert the sample tubing into the tting until it bottoms out.
9. Firmly tighten the nut so the ferrules crimp onto the sample tubing and make a seal.
Adjusting the Low Flow Setting
NOTE: Adjusting the low ow setting will cause a low ow alarm at the sample-draw
detector and a fail alarm at the controller. Be sure to put the controller into its
calibration program or disable external alarms before performing this test
The factory-set low ow setting is 0.6 SCFH (±0.2). To adjust the low ow setting:
1. Use the ow adjust potentiometer (VR1) to set the ow to 0.6 SCFH.
If the sample-draw detector goes into low ow alarm before you can adjust the ow
down to 0.6 SCFH, adjust the low ow potentiometer 1/4 turn clockwise, then
attempt to set the ow again. Repeat this step until you are able to adjust the ow to
0.6 SCFH.
NOTE: The low ow potentiometer is accessible through a circular cutout in the main
circuit board. The cutout is labeled PS1.
2. Slowly turn the low ow potentiometer counterclockwise just until the sample-draw
detector goes into low ow alarm.
3. Increase the ow using VR1 until the unit is out of low ow alarm.
4. Decrease the ow very slowly and verify that the low ow alarm is 0.6 SCFH (±0.2).
If the low ow alarm is set too low, turn the low ow potentiometer slightly
clockwise. Repeat steps 3 and 4 if necessary.
5. Use the ow adjust potentiometer (VR1) to set the ow to 1.5 SCFH.
6. Make sure the sample-draw detector’s Fail light is off.