65-2405RK-05 Combustible Gas Transmitter Operator’s Manual Part Number: 71-0180RK Revision: 0 Released: 2/16/11 RKI Instruments, Inc. www.rkiinstruments.
WARNING Read and understand this instruction manual before operating detector. Improper use of the detector could result in bodily harm or death. Periodic calibration and maintenance of the detector is essential for proper operation and correct readings. Please calibrate and maintain this detector regularly! Frequency of calibration depends upon the type of use you have and the sensor types.
Product Warranty RKI Instruments, Inc. warrants gas alarm equipment sold by us to be free from defects in materials, workmanship, and performance for a period of one year from date of shipment from RKI Instruments, Inc. Any parts found defective within that period will be repaired or replaced, at our option, free of charge.
Table of Contents Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 Specifications. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 Combustible Gas Detector. . . . . . . . . . . . . . . . . .
Overview This manual describes the 65-2405RK-05 combustible gas transmitter. This manual also describes how to install, start up, configure, maintain, and calibrate the transmitter when it is used with a gas monitoring controller. A parts list at the end of this manual lists replacement parts and accessories for the combustible gas transmitter. Specifications WARNING: Do not use this product in a manner not specified in this manual. Table 1 lists specifications for the combustible gas transmitter.
Description This section describes the components of the combustible gas transmitter. The transmitter is a 4 - 20 mA type detector head. It consists of the combustible gas detector, amplifier, and junction box.
Amplifier The amplifier converts the electrical output from the detector to a 4 to 20 mA signal that corresponds to the detection range and transmits the signal to a gas monitoring controller. A foam gasket that orients the amplifier and keeps it from rotating is installed on the bottom of the amplifier. A label on the amplifier indicates the detector drive current. This drive current is factory set and is dictated by the combustible gas to which the detector is calibrated. Consult RKI Instruments, Inc.
from a mounting surface and ensure that there is enough room to install a calibration cup on the detector during calibration. Installation This section describes procedures to mount the combustible gas transmitter in the monitoring environment and wire the transmitter to a controller. Mounting the Combustible Gas Transmitter 1. Select a mounting site that is representative of the monitoring environment. Consider the following when you select the mounting site.
Wiring the Combustible Gas Transmitter to a Controller WARNING: Always verify that the power to the controller is off before you make wiring connections. 1. Turn off the controller. 2. Turn off or unplug power to the controller. 3. Remove the junction box cover. 4. The detector leads are factory wired. Verify that the detector leads are wired to the amplifier’s detector terminal strip as shown in Figure 3. 5.
+ 24 VDC 4 - 20 mA In (S) - (DC Ground) Cable Shield Controller or Recording Device J-Box S R W LEL G B PWR / SIG Amplifier (Transmitter) Black Green White Red Detector Wires LEL Detector Figure 3: Wiring the Combustible Gas Transmitter to a Controller 12. If shielded cable is used, connect the cable’s drain wire to an available chassis (earth) ground at the controller. RKI controllers typically have a ground stud that can be used to ground the cable’s drain wire.
Start Up This section describes procedures to start up the combustible gas transmitter and place the transmitter into normal operation. Introducing Incoming Power 1. Complete the installation procedures described earlier in this manual. 2. Verify that the power wiring to the controller is correct and secure. Refer to the controller operator’s manual. 3. Turn on power to the controller. 4. Turn on the controller. 5. Verify that the controller is on and operating properly.
Maintenance This section describes maintenance procedures. It includes preventive maintenance, troubleshooting, and component replacement procedures. Preventive Maintenance This section describes a preventive maintenance schedule to ensure the optimum performance of the combustible gas transmitter. It includes daily, monthly, and quarterly procedures. Daily Verify a display reading of 0 %LEL at the controller. Investigate significant changes in the display reading.
6. Use the following formula to determine the correct test points output for the test sample. Output (mV) = (calibrating sample/fullscale) X 400 + 100 For example, with a test sample of 50 %LEL and a fullscale setting of 100 %LEL, the correct output is 300 mV. 300 (mV) = (50/100) X 400 +100 7. Screw the calibration cup onto the bottom of the detector. 8. Screw the regulator into the calibration cylinder. 9. Use the calibration kit sample tubing to connect the regulator to the calibration cup.
Table 2:Troubleshooting the Combustible Gas Transmitter Condition Symptom(s) Probable Causes Recommended Action Fail Condition • Controller indicates a fail condition. • The transmitter wiring is disconnected or misconnected. • The transmitter’s zero reading is low enough to cause a fail condition. • The transmitter is malfunctioning. 1. Verify that the transmitter wiring is correct and secure. 2. Calibrate the transmitter. 3. If the fail condition continues, replace the detector. 4.
8. Connect the detector leads to the detector terminal strip as shown in Table 3 below and Figure 3 on page 6 of this manual. Table 3:Reconnecting the Combustible Gas Detector to the Amplifier 9. Detector Lead Detector Terminal Strip Red LEL “R” White LEL “W” Green LEL “G” Black LEL “B” Reinstall the detector terminal strip into its socket. 10. Reinstall the junction box cover. 11. Turn on or plug in power to the controller. 12. Turn on the controller and place it into normal operation.
Table 4:Reconnecting the Amplifier to the Controller Amplifier Controller Terminal Strip Controller Transmitter Terminal Strip (typical) PWR/SIG “-” - (DC -) PWR/SIG “S” S (4 - 20 mA In) PWR/SIG “+” + 24V Table 5:Reconnecting the ppm Detector to the Amplifier Amplifier Detector Terminal Strip Detector Lead DETECTOR “R” RED DETECTOR “W” WHT DETECTOR “G” GREEN DETECTOR “B” BLK NOTE: When a transmitter is first powered up with a new amplifier, the initial output may be either high or below ze
Calibration This section describes how to calibrate the combustible gas transmitter. It includes procedures to prepare for calibration, set the zero reading, set the response reading, and return to normal operation. WARNING: The controller is not an active gas monitoring device during the calibration procedure. NOTE: The following procedure assumes the use of a calibration kit which includes a calibration gas cylinder, a 0.
4. Allow the gas to flow for one minute and verify a reading of 100 mV (±2 mV). If necessary, use the zero pot on the amplifier to adjust the reading to 100 mV (±2 mV). 5. Turn the regulator’s on/off knob clockwise to close the regulator. 6. Unscrew the regulator from the zero air calibration cylinder. 7. Leave the sample tubing connected to the regulator and the calibration cup. Setting the Response Reading 1. Screw the regulator into the calibration cylinder.
Parts List Table 6 lists replacement parts and accessories for the combustible gas transmitter. Table 6:Parts List Part Number Description 06-1248RK Sample tubing (3/16 in. x 5/16 in.; specify length when ordering) 18-0001RK Reducer, 3/4 in. NPT x 1/2 in.