USER MANUAL – BEDIENUNGSANLEITUNG RND 320-KA3305P
Safety Symbols This chapter contains important safety instructions that you must follow when operating the RND 320-KA3305P model and when keeping it in storage. Read the following before any operation to insure your safety and to keep the best condition for the RND 320-KA3305P model. Safety Symbols These safety symbols may appear in this manual or on the series..
RND 320-KA3305P USER MANUAL Safety Instruction Safety Guidelines • Do not block or obstruct the cooling fan vent opening. • Avoid severe impacts or rough handling that leads to damage. • Do not discharge static electricity. • Do not disassemble unless you are qualified as service personnel. AC Input • AC Inut Voltage: 230 V, 50 Hz • Connect the protective grounding conductor of the AC power cord to an earth ground, to avoid electrical shock.
Fuse MODEL RND 320-KA3305P 220/230 V T5 A/250 V(20 x 5 mm) • To ensure fire protection, replace the fuse only with the specified type and rating. • Disconnect the power cord before fuse replacement. • Make sure the cause of fuse blowout is fixed before fuse replacement.
RND 320-KA3305P USER MANUAL Front Panel Overview 5
Function Introduction 1 The Operation and Outputs of Voltage and Current 1. Press the button CH1 and then the CH1 display indicator blinks; press CH1 again and then some voltage digit on the channel 1 blinks. At this time, the voltage value can be set by adjusting the voltage adjustment knob. Then press the button CH1 again while the voltage digit is blinking to switch into the current digit blinking, when the current value can be set by adjusting the current adjustment knob.
RND 320-KA3305P USER MANUAL 2 The Series and Parallel Operation 1. THE SERIES OPERATION: Press and hold the button CH1 for 3 seconds to be in the series mode. When the SER indicator on the display lights on, it means the power supply is in the series mode now. At this mode, CH2 is the master operation while CH1 is the slave operation, when the CH1 operation is shielded. At this time, press the button ON/OFF and then the output can be turned on or off.
3 Recall to output In any state, just press the buttons M1 M5 and then the according memories can be recalled. 4 The Operation of Overcurrent Protection Press and hold the button “OCP” for 3 seconds to enter the mode of OCP setting, when the indicator “OCP SET” lights on. And the current values on both CH1 and CH2 display the OCP values accordingly. By adjusting the current knob, the OCP value can be changed. Press and hold the button “OCP” again for 3 seconds to exit.
RND 320-KA3305P USER MANUAL 6 Keyboard Lock Press and hold the voltage adjustment knob for 3 seconds, and then the front panel will be locked; press and hold it again for 3 seconds, and then it will be unlocked. 7 Beep ON/OFF Press and hold the current adjustment knob for 3 seconds, and then the beep will be turned off. Press and hold it again for 3 seconds, it will be turned on.
Remote control COM setting Set up the COM port inside the PC according to the following list. • Baud rate: 9600 • Parity bit: None • Data bit: 8 • Stop bit: 1 • Data flow control: None The Definition of Interface RS232 RND 320-KA3305P POWER SUPPLY Functionality check Run this query command via the terminal application such as MTTTY (Multi-threaded TTY). *DIN? This should return the identification information: Manufacturer, model name, software version.
RND 320-KA3305P USER MANUAL RND 320-KA3305P model Remote Control Syntax V2.0 Command format: VSET: 1. VSET: command header 2. X: output channel 3. : separator 4. NR2: parameter Command Details: 1. LOCK Description: LOCK or UNLOCK the front panel Example: LOCK1 LOCK the front panel Example: LOCK0 UNLOCK the front panel 2. ISET: Description: Sets the output current. Example: ISET1:2.225 Sets the CH1 output current to 2.225A 3. ISET? Description: Returns the output current setting.
5. VSET? Description: Returns the output voltage setting. Example: VSET1? Returns the CH1 voltage setting 6. IOUT? Description: Returns the actual output current. Example: IOUT1? Returns the CH1 output current 7. VOUT? Description: Returns the actual output voltage. Example: VOUT1? Returns the CH1 output voltage 8. OUT Description: Turns on or off the output. Boolean: 0 OFF,1 ON Example: OUT1 / Turns on the output 9. BEEP Description: Turns on or off the beep.
RND 320-KA3305P USER MANUAL 11. *IDN? Description: Returns the KA3005P identification. Example: * IDN? Contents KORAD KD3005P V2.0 (Manufacturer, model name,). 12. RCL Description Recalls a panel setting. 13. SAV Description: Stores the panel setting. NR1 1 5: Memory number 1 to 5 Example: SAV1 Stores the panel setting in memory number 1 14. TRACK Description: Sets the output of the power supply working on indepent or tracking mode NR1: 0=INDEP, 1 SER, 2=PARA Example: TRACK1 15.
Specifications Note: The specifications below are tested under the conditions of temperation 25°C to -5 °C and the warm-up for 20 minutes. RND 320-KA3305P- Model VOLTAGE RANGE CURRENT RANGE 0 - 30 V 0- 5 A LOAD REGULATION VOLTAGE RANGE CURRENT RANGE ≤ 0.01% +5 mV ≤ 0.1% +10 mA LINE REGULATION VOLTAGE RANGE CURRENT RANGE ≤ 0.01% +3 mV ≤ 0.1% +3 mA SETUP RESOLUTION VOLTAGE RANGE CURRENT RANGE 10 mV 1 mA SETUP ACCURACY ( 25°C + -5°C) VOLTAGE RANGE CURRENT RANGE ≤ 0.5% +20 mV ≤ 0.
RND 320-KA3305P USER MANUAL REACTION TIME VOLTAGE RANGE CURRENT RANGE ≤ 100 mS ≤ 100 mS (10% Rated load) LOAD REGULATION OF PARALLEL VOLTAGE RANGE ≤ 0.1% +0.1 V LOAD REGULATION OF SERIES VOLTAGE RANGE ≤ 0.1% +0.1 V CH3 SPECIFICATIONS(ONLY FOR 72-10505) VOLTAGE RANGE CURRENT RANGE VOLTAGE ACCURACY LOAD REGULATION 5V 3A ±50MV ±50MV ACCESSORIES SUPPLIED USER MANUAL 1 PC, POWER CORD 1PC WEIGHT AND DIMENSIONS(MM) 252(W)*135(H)*370(D), 9.
Deutsch Sicherheitssymbole Diese Sicherheitssymbole können in diesem Handbuch oder in der Serie erscheinen.
RND 320-KA3305P — BEDIENUNGSANLEITUNG Inhalt 1. Beschreibung und Funktion........................................................................... 18 2. Bestimmungsgemäßer Einsatz...................................................................... 18 3. Sicherheits-, Service- und Betriebshinweise................................................. 18 4. Bedien-, Anschluss- und Anzeigeelemente................................................... 19 5. Inbetriebnahme und Betrieb......................
1. Beschreibung und Funktion Das RND 320-KA3305D/P ist ein hochwertiges Dreifach-Labornetzgerät für den Konstantspannungs- und Konstantstrombetrieb. Es verfügt über eine fein einstellbare Stromund Spannungseinstellung, getrennte Anzeigen für Ausgangsstrom/ausgangsspannung, 5 Speicher für schnell abrufbare Strom-/Spannungseinstellungen sowie Schutzeinrich tungen gegen Überspannung/Überstrom.
RND 320-KA3305P — BEDIENUNGSANLEITUNG • Das Gerät ist kein Spielzeug, es gehört nicht in Kinderhände. Es enthält heiße, kleine und verschluckbare Teile sowie Kabel, die Strangulierungsgefahren bergen. • Lassen Sie Verpackungsmaterialien nicht achtlos liegen, diese können für spielende Kinder zu einer Gefahr werden.
Zustandsanzeige Ausgänge Netzschalter Konstantstrombetrieb Konstantspannungsbetrieb Speicher-Aktivanzeige 5 Lastanschluss Kanal 1 Lastanschluss Kanal 3 Erdungsanschluss Lastanschluss Kanal 2 Spannungseinstellung Anzeige Serien-/Parallelbetrieb Ausgang ein/aus Stromeinstellung Überstromschutz Überspannungsschutz Zustandsanzeige Überspannungsschutz Zustandsanzeige Überstromschutz Speicher 1...
RND 320-KA3305P — BEDIENUNGSANLEITUNG 5. Inbetriebnahme und Betrieb 5.1 Aufstellung • Stellen Sie das Gerät so auf, dass ringsum eine gute Luftzirkulation gewährleistet ist und es nicht längere Zeit eine Wärmequelle (Heizung, direkte Sonnenbestrahlung) ausgesetzt ist. 5.2 Inbetriebnahme • Trennen Sie eine eventuell angeschlossene Last vom Netzgerät. • Schließen Sie das Gerät mit dem mitgelieferten Netzkabel an eine Schutzkontaktsteckdose an. • Schalten Sie das Netzgerät mit dem Netzschalter ein.
5.5 Last an- und ausschalten • Mit der Taste „Off/On” schalten Sie den Lastanschluss an das Netzgerät an oder ab. Im eingeschalteten Zustand leuchtet die Anzeige „ON”.. 5.6 Bedienelemente blockieren • Zur Vermeidung eines versehentlichen Verstellens oder unbefugter Bedienung drücken Sie den Spannungseinsteller für ca. 3 s, bis ein Quittungston ertönt. • In gleicher Weise entsperren Sie das Bedienfeld wieder. 5.
RND 320-KA3305P — BEDIENUNGSANLEITUNG 5.10 Serien-/Parallelschaltung • Für die Serienschaltung der beiden regelbaren Teilnetzteile drücken Sie die Kanaltaste „CH1” so lange (ca. 3 s), bis die „SER”-Anzeige leuchtet. • Jetzt sind beide Kanäle in Serie geschaltet, wobei Kanal 2 (CH2) nun als Master und CH1 als Slave fungiert. Alle Einstellungen werden also für Kanal 2 vorgenommen, Kanal 1 folgt diesen.
7. Wartung, Lagerung und Pflege • Lagern Sie das Gerät sauber, kühl und trocken. • Reinigen Sie das Gerät nur mit einem trockenen Leinentuch. Nicht zu stark auf die Frontplatte drücken! Bei stärkeren Verschmutzungen kann das Reinigungstuch leicht mit Wasser angefeuchtet sein.