User Manual
16910 CoolTech ID Plus Refrigerant Identifier
17
Post-Operational Procedure
1. Unplug the instrument power cord.
2. Disconnect the sample hose from the inlet port of the instrument. Inspect the hose for
signs of wear, such as cracking, fraying, or kinks. Verify the hose is not obstructed, and
that no oil is present inside the hose. Store the inspected hose in the storage case. If the
hose shows signs of wear, obstruction, or oil, it MUST BE REPLACED BEFORE
USING THE INSTRUMENT AGAIN.
3. Disconnect the purge vent hose, if it is connected, and clean and stow it in the storage
case. Replace the cap onto the purge vent.
WARNING
ALWAYS replace the cap onto the purge vent port. Failure to replace the cap
can result in excessive refrigerant loss during the next use of the instrument,
which is a health hazard.
4. Inspect the sample filter for signs of red spots or discoloration on the white outside
diameter of the filter element. If any red spots or discoloration are noticed, REPLACE
THE FILTER BEFORE USING THE INSTRUMENT AGAIN.
CAUTION
When red spots or discoloration begin to appear on the white outside diameter
of the filter element, THE FILTER MUST BE REPLACED. Failure to correctly
maintain the sample filter may result in severe instrument damage that is not
covered under warranty repairs.
5. Using a moist cloth, clean the instrument of accumulated dirt, oil, grime, etc. Do NOT
USE SOLVENTS OR LIQUIDS TO CLEAN THIS INSTRUMENT. Stow the
instrument in the storage case.
Operating Procedures