Meter Maintenance and Troubleshooting
9
Not enough blood or control
solution was drawn into the
test strip for measurement or
was applied after the test had
started.
Discard the test strip and
repeat the blood glucose or
control test.
Blood or control solution was
applied to the test strip before
Apply drop appeared.
Discard the test strip and
repeat the blood glucose or
control test.
An electronic error occurred.
Remove the batteries, press
and hold the Power/Set/OK
button for at least 2 seconds,
and reinsert the batteries.
Perform a blood glucose or
control test.
The temperature is above or below the proper
range for the system.
Refer to the test strip package insert for system
operating conditions. Move to an area with the
appropriate conditions and repeat the blood
glucose or control test. Do not artificially heat or
cool the meter.
The batteries are out of power.
Change the batteries now. If the message
reappears after the batteries have been replaced,
remove the batteries, press and hold the Power/
Set/OK button for at least 2 seconds, then
reinsert the batteries.