User Manual

13
263
13.2 Cleaning and Disinfecting the Meter
To clean and disinfect without damaging the meter, follow these
procedures carefully.
When to Clean and Disinfect the Meter
j Clean the meter to remove visible dirt or other material prior to
disinfecting.
j Clean and disinfect the meter at least once per week, or when
blood is present on the surface of the meter.
j Clean and disinfect the meter before allowing anyone else to
handle the meter. Do not allow anyone else to use the meter
on themselves for testing purposes.
NOTE
Using cleaning and disinfecting products could result in
damage to the meter. If you notice any of the following
signs of deterioration after cleaning and disinfecting your
meter, stop using your meter and contact the ACCU‑CHEK
Customer Care Center at 1‑800‑688‑4578: residue around
buttons, clouding of display, button/touchscreen
malfunction, out‑of‑range results.
52297_07167776001.indb 263 10/20/14 11:19 AM