User Manual

Screen
Message
Possible Causes Possible Solutions
Insulin
sensitivity not
valid
The insulin sensitivity is outside of the
acceptable meter range.
Check your entries and contact your healthcare professional to
determine the appropriate settings.
Invalid bolus
advice times
The acting time value is less than the offset
time value.
The acting time value must be set equal to or greater than the offset
time value. Reset the acting time value or revise the offset time
value.
Invalid setting
with the text
Hypo value
must be less
than time block
target ranges.
The hypo warning limit value must be less
than all of your target ranges in the time
block settings.
Either reset the hypo warning limit so it is below the target ranges of
your time blocks or revise the target ranges for the time blocks. See
the Warning Limits: Hypo, Hyper section in the Changing Meter
Settings chapter in this user’s manual.