User Manual

Table Of Contents
Testing a Blood Sample • 4
115
STAT tests STAT tests are a limited number of tests that can be per-
formed in emergency situations, see page 88. If the meter
is configured to perform STAT tests, you have the option
of performing a test by choosing STAT even though a QC
Lockout is in place.
1 To perform the measurement without performing a
quality control test, touch Patient Test:
2 Touch the button to select the parameter you want
to use.
3 Touch to confirm the number of remaining
STAT tests.
4 Perform the test.
When a STAT test is performed, this information is stored
with the result. The number of STAT tests allowed is
reduced by 1. After all pending quality control tests are
performed, the specified number of STAT tests is availa-
ble again in case of a new lockout.
Main Menu
09:15 am
05/29/2015
Control Test
Review Results
Setup
Patient T
est
Logout
QC!
Test
09:15 am
05/29/2015
PT
QC status
LockedaPTT
3
-
-
Test
09:15 am
05/29/2015
aPTT
PT
STAT Info
3 STAT tests remaining
before lockout.
I-011
The default setting is three for the number of STAT
tests allowed when in QC Lockout.
Test
09:15 am
05/29/2015
P-ID: PID01
Code: 184
2.5
INR
Smith, Henry
Op: Operator 1
Par: PT
ST
AT Test
21
%Q
05/29/2015 09:14 am