12 Apply a single drop of the sample directly from the
dropper to the semicircular, transparent sample
application area on top of the test strip. Do not add
more of the sample.
You hear a beep when you have applied enough of the
sample (provided the beeper is enabled). The dropper
icon disappears and the test starts.
The result of the proficiency test is displayed, the com-
ment Proficiency is automatically added. If you want to
add further comments, touch . The test result is auto-
matically saved to memory.
After the test result is displayed, touch . You will be
prompted to remove the strip.
13 Remove the test strip from the meter.
14 Power the meter off.
15 Clean the meter if this becomes necessary (see
Chapter 8, Maintenance and Care).
Profi ciency
09:15 am
05/29/2015
QC: 202
Code: 101 Par: aPTT
Profi ciency
09:15 am
05/29/2015
QC: 202
Code: 101
SEC
06/03/2015 11:30 am
Par: aPTT
Profi ciency
51.6
Dispose of used test strips from proficiency testing
in
accordance with the disposal policy of your facility
.