QC Range Liquid quality control ranges can be customized to
comply with local guidelines. The QC Range function
enables you to narrow the default range individually for
each test parameter.
1 From the Main Menu, touch Setup to open the
meter settings.
2 From the Setup Menu, touch QC Settings.
3 From the QC Settings menu, touch QC Range.
4 From the QC Range menu, touch the button of the
test parameter you wish to set.
You may select from the following options:
Default Range (Range provided in the code chip
is selected and displayed. No target value is dis-
played.)
Custom Range (The allowed percentage deviation
from target value can now be customized. Addi-
tionally, you can choose between displaying the
target value along with a control test result or not
displaying the target value with the control test
result.)
5 Touch the button with the setting of choice. Your
selection is now highlighted.