Operator`s manual
Maintenance and Care
122
Cleaning the base unit
■ Unplug the base unit before cleaning. 
■ Wipe the surfaces with a soft cloth slightly 
dampened (not wet).
■ Avoid wiping the electrical connectors on the back 
of the base unit.
■ Dry the base unit thoroughly after cleaning. 
Visually verify that no solution is seen in the base 
unit connectors at the completion of cleaning. If 
cleaning fluid is allowed to collect in any connector, 
severe damage can occur to the meter and the base 
unit.
■ If using commercially available pre-moistened 
cleaning cloths, squeeze off excess cleaning 
solution or blot on a dry paper towel to remove any 
excess cleaning solution before cleaning the 
surface of the base unit.
■ If cleaning solution does get on the connectors or 
pools in the base unit, unplug the base unit, then 
dry the components with a dry cloth or gauze pad 
before returning the meter to the base unit.
■ Plug in the base unit.
NOTICE
Ensure meter/base unit (including connectors) are thor-
oughly dried after cleaning or disinfecting. A flashing 
LED (red) on the base unit indicates a fault condition.
NOTICE
Do not spray the base unit directly with solutions as this 
could cause the solution to enter the case and damage 
the electronic components.