Operator`s manual
Troubleshooting
131
11 Troubleshooting
The Accu-Chek Inform II meter continually checks its sys-
tems for unexpected and unwanted conditions. 
A troubleshooting table follows that will help you when 
the system is not performing as expected. Most concerns 
can be resolved quickly by referring to this table for help. 
Take the following steps when an unexpected condition 
arises:
■ Find the displayed message or condition in the 
Troubleshooting Table.
■ Take the action suggested under the column 
header Possible Solution. 
If further assistance is required, call your local Roche 
representative.
Errors and unusual behavior 
without error messages
Some conditions may arise that have no associated error 
message. Refer to the following table, if such a condition 
arises on the Accu-Chek Inform II system.
Display/symptom Possible solution
No message or unusual behavior
Meter display does not turn on – Wait 10 seconds and try turning on the unit again. 
– Place the meter in the base unit and confirm that it is 
charging.
– Check that the battery pack is correctly installed and 
connected.
Follow the instructions in Chapter 9 to avoid data loss when 
replacing the battery pack.
Meter displays an unexpected result Refer to the package insert for the test strips.