Operator`s manual
Troubleshooting
136
Meter reset A meter reset can be performed to correct a number of 
unspecified errors (e.g., “frozen” screen, etc.). Use the fol-
lowing steps to perform a meter reset. 
1 Place the meter face down on a level surface.
2 Press the reset button in the middle of the battery 
pack using a tool such as a screwdriver or a paper 
clip.
3 Turn over the meter.
■ The Roche logo is displayed. 
If the Roche logo does not appear within 60 
seconds, place the meter on a base unit for a 
minimum of 15 minutes to recharge the battery.
■ Within a short period of time, the Versions screen 
will appear, displaying the version numbers of the 
meter components.
4 The screens for entering the date and the time 
appear if date and time need to be set. After you 
have entered the correct information, confirm each 
screen with  .
■ Finally, the Power Up screen and the Main Menu 
are displayed as usual. 
Depending on the configuration of your system, the 
meter may have to be docked in the base unit for 
synchronization of the meter's date/time with the 
date/time of your facility.