Operator`s manual
Appendix
145
Patient ID validation mode 0: none
1: length
2: list
3: list allowing entry if not 
on list
4: length if numeric
0YY
Patient ID length
(Allowed characters: a-z, 0-9, 
“.” [period], “-” [hyphen])
1-20 characters N/A N Y
Patient name length 0-25 characters N/A N N
QC algorithm
QC Algorithm 0: None (always OK)
1: Last result OK
2: Time of day (DMS only)
3: Shift (DMS only)
4: Hours
5: Strip count
6: Time of day rotate 
(DMS only)
7: Shift rotate (DMS only)
8: Hours rotate (DMS 
only)
9: Strip count rotate 
(DMS only)
0 (Y) Y
QC hours (Number of hours between 
required control measurements if QC 
algorithm is set to “Hours” or “hours 
rotate”)
0 – 9,999hours 24 Y Y
QC shift length (a PASS control remains 
valid for twice this amount in time after 
shift start if the QC algorithm is set to 
“Shift” or “Shift rotate”)
1, 2, 3, 4, 6, 8, 12, 24 hours 8 N Y
QC shift start (time of the shift start if the 
QC algorithm is set to “Shift” or “Shift 
rotate”)
0 – 23 hours 6 N Y
QC strip count (number of tests a PASS 
control remains valid)
1 – 999 50 Y Y
QC time of day 6 POCT1-A data fields 
00:00 – 23:59
06:00
09:00
12:00
15:00
18:00
22:00
NY
QC time of day set 6 POCT1-A data fields 
0 or 1
1 0 0 0 0 0 N Y
Subject/Attribute Range Default Device DMS