Radiofrequency radiation 
exposure information
No operating limitations apply to meters without an RF 
card or to meters with an RF card that are not docked.
This equipment complies with FCC radiation exposure 
limits set forth for an uncontrolled environment. This 
equipment should be installed and operated with mini-
mum distance of 20 cm (8 inches) between the radiator 
and your body.
This transmitter must not be co-located or operated in 
conjunction with any other antenna or transmitter. 
Changes or modifications made to this equipment not 
expressly approved by Roche Diagnostics may void the 
FCC authorization to operate this equipment.
Glossary: 
■ “FCC” stands for “Federal Communications 
Commission” (USA).
■ “RSS” stands for “Radio Standards Specification” 
(Canada)
Note: The following table provides an overview of the 
classifications for electromagnetic compliance (EMC, 
according to CISPR 11) for different combinations of 
Accu-Chek Inform II system components. Refer to the 
explanatory texts following the table for more 
information on these classifications.
Classification acc. to CISPR 11 Meter with RF, docked, 
serial number 
<UU11030000
Meter with RF, docked, 
serial number 
≥UU11030000
Accu-Chek Inform II Base Unit, serial 
number <UU41030000 and connected 
to network / a PC via a cable
Class A Not allowed
Accu-Chek Inform II Base Unit, serial 
number ≥ UU41030000
Class B Class B
Note: The following Classification B applies to all 
hardware combinations except those specified on 
page 19.