Note on systems with older 
hardware
The Accu-Chek Inform II system complies with the emis-
sion and immunity requirements described in IEC 61326-
2-6:2006. It has been designed and tested to CISPR 11 
Class A. In a domestic environment it may cause radio 
interference, in which case, you may need to take meas-
ures to mitigate the interference. The electromagnetic 
environment in which the Accu-Chek Inform II device is 
to be operated should be thoroughly evaluated prior to 
operation of the device.
WARNING
Possibility of Electromagnetic Interference
Under certain conditions, a specific combination of hard-
ware equipment may emit electromagnetic radiation of 
such high energy that it can interfere with the proper 
functioning of other electronic or medical equipment. 
Therefore it is not allowed to dock an Accu-Chek Inform 
II meter with a serial number above or equal to 
UU11030000, which is equipped with an RF card, into an 
Accu-Chek Inform II Base Unit with a serial number 
below UU41030000 if this Accu-Chek Inform II Base Unit 
is connected to a network directly or through a PC via a 
cable.
Note: The following classification applies if an 
Accu-Chek Inform II meter (with RF card) with a serial 
number below UU11030000 is docked in an 
Accu-Chek Inform II Base Unit with a serial number 
below UU41030000 if this Accu-Chek Inform II Base 
Unit is connected to a network directly or through a 
PC via a cable.