Operator`s manual
Patient Glucose Testing
33
3 Patient Glucose Testing 
Information regarding blood glucose testing
Preparing to test The following requirements must be met before you can 
perform a test:
■ The Accu-Chek Performa or Accu-Chek Inform II 
test strips are available. 
■ At least one code file for test strips must be stored 
in the meter and match the lot number of the test 
strips used (see Chapter 6).
■ An operator ID must be entered (with password, if 
required), if the meter is configured for login.
■ Glucose control tests specified in the system 
configuration must be run and completed 
successfully before running patient tests. 
Information whether glucose control tests are 
necessary is displayed in the Power Up screen.
■ If configured, the meter may require downloading 
stored data to the data management system within 
defined time intervals. If such a download (either 
via WLAN or by docking the meter in a base unit) 
does not happen within the specified time, the 
meter is locked (Download Lockout) and cannot be 
used for testing. 
If a glucose control test is shown as required, patient glu-
cose testing cannot be performed until the controls are 
run successfully. For emergencies, STAT tests can be 
configured in the meter. This option allows the meter to 
perform a limited number of patient glucose tests, if cir-
cumstances require, despite the meter being in QC Lock-
out or Download Lockout (see page 57).
Contrast
Power Up 12:48
pm
03/16/10
QC Due: Immediately
■ Observe the applicable regulations and directives 
for hygiene and safety when collecting blood 
samples.
■ Observe the applicable regulations and directives 
for disposing of potentially infectious samples and 
materials.