Operator`s manual
Glucose Control Testing
48
Glucose control testing intervals Intervals between running glucose control tests are 
determined by your facility. These intervals are entered 
when the system is configured. At the end of the speci-
fied interval (or after a specific event such as starting to 
test with a new test strip lot), a warning is displayed when 
the meter is switched on and when the Glucose Test 
function is selected. 
Glucose control tests should be run in the following cir-
cumstances:
■ Before using the meter for patient testing the first 
time
■ At the glucose control intervals established by your 
facility
■ When using a new test strip vial for the first time
■ When using a new test strip lot for the first time 
(and as a result a new test strip code)
■ If a test strip vial was left open
■ If questionable test results are displayed repeatedly
■ If you wish to test the performance of the system
In addition, the following events can be specified during 
setup as a reason for a glucose control test:
■ If a previous control test is out of range
■ If glucose control tests were not run at the proper 
intervals
If a glucose control test is required (as shown in the illus-
tration on the left), you will not be able to test blood glu-
cose until the controls have been run successfully. For 
emergencies, STAT tests can be configured in the meter. 
This option allows the meter to perform a limited number 
of blood glucose tests, if circumstances require, despite 
the meter being in QC Lockout (see page 57).
Contrast
Power Up 12:48
pm
03/16/10
QC Due: Immediately