Applying the control solution Once the meter has detected the test strip, you are 
prompted to apply the control solution. 
1 Wait until the flashing drop appears in the display 
before applying the control solution. 
2 Apply a drop of glucose control solution to the 
front edge of the test strip. Do not apply the 
control solution to the top of the strip.
The control solution is pulled into the test strip by 
capillary action.
Once sufficient control solution has been detected, the 
meter beeps and the measurement begins.
Control Test 12:48pm
03/16/10
Control L2 (Hi) 123456
Strip Lot 545794