Operator`s manual
Proficiency Testing
90
Applying a proficiency sample Once the meter has detected the test strip, you are 
prompted to apply the proficiency sample. 
1 Wait until the flashing drop appears in the display 
before applying the sample. 
2 Apply a drop of sample to the front edge of the test 
strip. Do not apply the sample to the top of the strip.
The sample is pulled into the test strip area by 
capillary action.
Once sufficient sample has been detected, the meter 
beeps and the measurement begins.
Proficiency Test 12:48pm
03/16/10
Sample 123456789
Strip Lot 545794