User`s guide
HP 300-121 
Page 4 of 5 
Reviewed/Revised 06/26/13 
15. After the sample has been obtained, apply gentle pressure to the puncture with a clean gauze square 
or cotton ball site for several minutes. 
16. The measurement is complete when the result is displayed on the screen.  Depending upon how high 
or low the result is, it may appear in a numeric or non-numeric format.  See Reporting Results section 
below for interpretation of each result format. 
17. Remove the test strip and dispose of it in plain trash. 
18. Touch    to  enter  up  to  three  appropriate  comment(s)  as  required in  the  “General Policies” 
section of this manual. 
19. Touch the   button to confirm the result and send the result from the meter wirelessly or place the 
meter in the base unit to send the result and record the result into the electronic data management 
system. The base unit also charges the meter. 
20. Documentation of the result in the patient’s electronic medical record occurs automatically, as long as 
the patient’s FIN number is entered correctly. 
21. Follow up on any results that exceed critical or reportable limits according to policy. 
22. Clean and disinfect as necessary with Clorox wipes. 
23. Error messages: 
  If the error message “Strip Defect Error” appears on the display, the test strip may be defective or 
the blood glucose result may be extremely low and below the meter’s measurement range.  Refer 
to the test strip package insert, perform a quality control test using a new test strip, review proper 
testing procedure, and repeat the blood glucose test, or follow your facility’s testing policy. 
  If the meter displays “Type Bad Dose,” there may be insufficient amount of blood on the test strip. 
Repeat the test using a new test strip, ensuring proper sample application, or refer to the test strip 
package insert. 
  Guidance for Interpreting On-Screen Message and Error Codes: 
All  error  messages  displayed  by the  system have  a  letter  identifying the  message  type, a 
number and a description of the error to help the operator take action to resolve the problem. 
The different message types are in the table below. 
E 
Identifies  the  notification  as  an  Error.    The  information  notifies  the 
operator that an error has occurred. 
W 
Identifies the notification as a Warning.  The information does not block 
the operator from  continuing, but rather  gives the operator  information 
that may suggest an alternate workflow is required. 
I 
Identifies  the  notification  as  Informational  only.    Informational 
notifications present the operator with contextual information, and allow 
the operator to proceed after confirming the notification. 
D 
Identifies a Decision point.   Decision notifications  provide  the  operator 
with a choice based on contextual information. 
Linearity Testing Procedure 
Linearity testing, also known as calibration verification, is a means of testing the ACCU-CHEK Inform II 
system’s ability to give correct readings through a broad range of glucose levels.  It is used as a periodic 
check on system performance.  Linearity testing is performed on 10% of meters during new  instrument 
implementation, then on all subsequent meters or as part of troubleshooting related to questionable meter 
performance. 
Proficiency Testing Procedure 
Participation  in  proficiency  testing  is  part  of  the  quality  assurance  program  to  ensure  proper  on-going 
performance of the ACCU-CHEK Inform II system and the competence of staff performing tests.  At least 
10% of meters are tested during each proficiency cycle by staff routinely using the device.  Results are 
submitted to CAP without consulting with any other survey subscribers.  The survey report is reviewed by 










