Technical information

Understanding Control Test Results
20
Chapter 2: Control Testing ACCU-CHEK Customer Care Service Center 1-800-858-8072
The label on your test strip container shows the acceptable
ranges for low and high test solutions. Your control test
results should be within the listed range. Make sure you
compare the result to the range for the level of control
solution you used (low or high). The low level control
solution bottle has a blue cap; the high level solution bottle
has a red cap.
When a control test result is within the acceptable range,
your test strips and meter are working properly.
Range (mg/dL)
LOW 42–72
HIGH 290–328
Example
Example