4342
Error: Code Chip
The code chip is missing,
not properly inserted, or
damaged.
Solution:
Check to see if you have
the correct code chip
properly inserted in the meter.
For more information, see
the Code Chip section of
this manual.
If the code chip is damaged,
call Roche Diagnostics
Technical Service Center
at 1-800-428-4674.
Error: Test Strip Expired
The test strip has expired.
Solution:
Check the meter’s date
setting. If it is not correct, set
the correct date. For more
information, see the Meter
Setup section of this manual.
If the date is correct, turn the
meter off and remove the
code chip and the test strip.
Then use the code chip and a
test strip from a new box of
test strips.
Error: Test Strip Unusable
The test strip is unusable.
Solution:
Turn the meter off, remove
the test strip, and then
re-insert it. If the error
message reappears, discard
the unusable test strip and
use a new one.
Error: Test Strip Guide Cover
The test strip guide cover is
not properly closed.
Solution:
Close the test strip guide
cover.