User manual

4544
Error: Blood Application
Error applying blood to the
test strip.
Solution:
Turn the meter off and remove
the test strip.
Repeat the test using a new
test strip and blood taken
from a new fingerstick from a
different finger.
Error: Test Strip Interference
The test strip was touched or
removed during the test.
Solution:
Turn the meter off and
remove the test strip. Repeat
the test using a new test
strip and blood taken from a
new fingerstick from a
different finger.
Do not touch or remove
the test strip when a test is
in progress.
Error: Quality Control Failure
The test strip failed the
internal quality control check.
The test strip is unusable.
Solution:
Turn the meter off and
remove the test strip. Repeat
the test using a new test
strip and blood taken from a
new fingerstick from a
different finger.
Error: Time Exceeded
You did not apply blood to
the test strip within 2 minutes
after the blood drop symbol
appeared.
Solution:
Turn the meter off and
remove the test strip. Repeat
the test using the same test
strip and blood taken from a
new fingerstick from a
different finger.
Error Messages You May See
After You Apply Blood