Manual

Publication GMSI10-UM008D-EN-P - August 2010
66
Inputs
Channels
Supported Thermocouple Types
Supported RTD Types
1 to 6 RTD or thermocouple transducer
signals, user configurable
XM-361 accepts RTD and thermocouple
inputs
XM-362 accepts thermocouple inputs only
100 ohm 2-wire & 3-wire Platinum
(alpha = 0.00385):
-40 to 660°C (-40 to 1220°F)
200 ohm 2-wire & 3-wire Platinum
(alpha = 0.00385):
-40 to 453°C (-40 to 847°F)
100 ohm 2-wire & 3-wire Platinum
(alpha = 0.003916):
-40 to 660°C (-40 to 1220°F)
200 ohm 2-wire & 3-wire Platinum
(alpha = 0.003916):
-40 to 443°C (-40 to 829°F)
250 ohm 2-wire & 3-wire Platinum
(alpha = 0.00392):
-40 to 389°C (-40 to 732°F)
100 ohm 2-wire & 3-wire Nickel
(alpha = 0.00618):
-40 to 180°C (-40 to 356°F)
120 ohm 2-wire & 3-wire Nickel:
(alpha = 0.00672):
-40 to 439°C (-40 to 822°F)
10 ohm 2-wire & 3-wire Copper:
(alpha = 0.00427):
-40 to 260°C (-40 to 500°F)
XM-361/362 Technical Specifications
Product Feature Specification
Type Range C° Range F°
B 0°C to 1810°C (32°F to 3290°F)
C 0°C to 1316°C (32°F to 2400°F)
E 5°C to 284°C (41°F to 543°F)
J 0°C to 364°C (32°F to 687°F)
K -40°C to 484°C (-40°F to 903°F)
N -40°C to 620°C (-40°F to 1148°F)
R -40°C to 1760°C (-40°F to 3200°F)
S -40°C to 1760°C (-40°F to 3200°F)
T -40°C to 379°C (-40°F to 714°F)