Specifications

Trace Settings R&S FSH
1145.5973.12 4.20 E-10
Operating sequence:
¾ Press the TRACE key.
¾ Press the TRACE MODE softkey.
The submenu for setting the trace mode opens.
¾ Using the cursor keys or the rotary knob, select the
trace mode you want and confirm with the ENTER
key or the TRACE MODE softkey.
The "Trace:" display at the center of the top of the
display shows the trace mode that has been selected.
If TRACE MODE AVERAGE is selected, the AVG COUNT entry field opens, which displays the set
number of averagings.
The following actions can be performed:
¾ Confirm the displayed number of averagings with
the TRACE softkey or the ENTER key.
¾ Using the numeric keypad, enter a new figure
between 2 and 999 for specifying the number of
averagings and confirm your entry with the TRACE
softkey or the ENTER key.
¾ Change the number of averagings by using the
rotary knob and confirm your entry with the TRACE
softkey or the ENTER key.
The R&S FSH averages the pixels of the trace across
the set number of averagings.
If the sweep is continuous, the R&S FSH then performs a sliding averaging. In the SINGLE SWEEP
mode, it performs exactly those sweeps defined with AVG COUNT and averages the traces. It then
stops the sweep and displays the averaged trace.
In the trace mode VIEW, the settings used for measuring the trace are displayed. This ensures that the
measurement conditions can be clearly specified in result documentation. In the status display
(STATUS key) it is indicated in brackets that the view mode is currently selected, e.g. Trace Mode:
Maximum Hold (View).