Specifications

R&S FSH Setting and Using the Measurement Functions
1145.5973.12 4.55 E-10
The peak detector is recommended. It ensures that the peak power of the vision carrier is correctly
measured.
The sweep time is coupled to span, resolution bandwidth and video bandwidth. Set the sweep time
in such a way that the filters can settle. Too short a sweep time distorts the measurement result.
¾ Press the SELECT MEASURE softkey.
The menu for selecting the reference measurement
opens.
¾ Using the rotary knob or the cursor keys, select
CW Tx and confirm your choice with the ENTER key
or the F1 softkey.
The R&S FSH starts the CW Tx reference
measurement. C/N-CW Tx appears in the top right-
hand corner of the display. Reference channel and
channel bandwidth are set in accordance with the last
CW Tx reference measurement.
In the CW Tx operating mode, the count marker is automatically activated. After each sweep, the
marker is set to the highest power in the reference channel. Marker power and marker frequency are
indicated at the top edge of the display. The marker power corresponds to the reference.
Manual reference mode:
As an alternative to the C/N reference measurement, the R&S FSH allows you to manually determine
the C/N reference.
¾ Press the REF MEASURE softkey.
¾ Using the rotary knob or the cursor keys, select MAN REFERENCE POWER / LEVEL and confirm
your choice with the ENTER key or the F4 softkey.
¾ Using the numeric keypad, enter the desired reference and terminate your entry with the appropriate
unit key or the ENTER key.
The R&S FSH indicates the selected reference at the bottom of the display.