Specifications

R&S FSH Setting and Using the Measurement Functions
1145.5973.12 4.63 E-10
Using the R&S FSH in receiver mode
(only available if option R&S FSH-K3 is installed)
The receiver mode (option R&S FSH-K3) is used for measuring levels on individual frequencies. In this
mode, the R&S FSH functions as a receiver and measures the level on a prescribed frequency.
In addition, it is possible to perform measurements on several frequencies with a graphical display of
the levels. Unlike the analyzer mode, in which the R&S FSH sweeps quasi-continuously over the
specified frequency range, in the receiver mode the instrument measures on specified discrete
frequencies using the selected measurement time per frequency.
Activating the receiver mode:
¾ Press the MEAS key.
¾ Press the MEASURE softkey.
The menu for measurement functions opens.
¾ Use the rotary knob or the cursor keys to select
RECEIVER from the menu and confirm with the
ENTER key or the MEASURE softkey.
The R&S FSH activates the receiver mode and measures the level on the set frequency.
Screen layout:
Main menu of
receiver mode
Frequency
Level
Analog level
Bandwidth
Measurement
time
Detector
Reference
level
Level scale
Trigger
The R&S FSH provides the most important setting parameters such as frequency, reference level,
measurement bandwidth, detector and measurement time in the main menu of the receiver mode.
However, the settings can also be made using the corresponding keys.