Specifications

R&S FSH Setting and Using the Measurement Functions
1145.5973.12 4.91 E-10
The R&S FSH then requests to terminate the output of
the tracking generators into 50 .
¾ Connect the generator output with the
50 termination.
¾ Press the F1 or F2 (CONTINUE) softkey.
During calibration, the R&S FSH outputs the message
"Calibrating LOAD, please wait".
When calibration is over, Transm (vect) is displayed
in the top right-hand corner to indicate that the
R&S FSH is vector-calibrated for transfer
measurement.
Connect the DUT between the tracking generator
output and the RF input. The R&S FSH displays the
magnitude of the transfer function of the DUT.