Specifications

Measuring with Transducer Factors R&S FSH
1145.5973.12 4.122 E-10
Frequency range of transducer:
If the set frequency range is wider than the span in which a transducer is defined, the R&S FSH
assumes the transducer values outside the defined range to be zero.
Data sets containing transducer factors:
The R&S FSH stores data sets together with any transducer factors that may have been active for the
measurement in question. When such a data set is recalled, the associated transducer factor(s) are
switched on as well. Transducer factors recalled as part of a data set do however not appear in the list
of transducer factors.