Specifications

R&S FSH Saving and Loading Instrument Settings and Measurement Results
1145.5973.12 4.123 E-10
Saving and Loading Instrument Settings and Measurement
Results
The R&S FSH’s settings and measurement results can be saved to the internal memory and recalled at
a later date. Using the R&S FSH View software package, these data sets can also be saved to a PC
from the R&S FSH or downloaded onto the R&S FSH from a PC.
Results and settings, including the measurement function, are always saved en bloc so that when the
results are recalled the measurement context is clear. The R&S FSH can store a maximum of 100 data
sets which are assigned a unique name.
Data sets for scalar transmission and reflection measurements can be stored along with their calibration
data. When such data sets are recalled, therefore, measurements can be performed without prior
calibration. Saving a data set with calibration data, however, requires twice as much memory space as
without it, i.e. a data set with calibration data takes up the space required for two data sets without
calibration data. This reduces the maximum number of data sets that can be stored by the number of
data sets stored with calibration data.
Storage of calibration data can be selected in the SETUP menu (see Chapter 2, section "Saving
Calibration Data“).
¾ Press the SAVE / PRINT key.
The R&S FSH opens the SAVE / PRINT menu where
the functions for saving, clearing and loading data sets
are displayed for selection.
A screenshot can also be output to a printer.
Saving results
¾ Press the SAVE softkey.
The R&S FSH opens a text box and prompts the user
to enter a name for the data set.
The Name entry box, which is highlighted in red, also
suggests a name for the data set (DATASET.000)
which can be accepted by pressing the ENTER key.
For the sake of simplicity, the R&S FSH also saves the
data set when the SAVE softkey under the suggested
name is pressed twice.
The remaining free memory locations
(Free Locations) are also displayed in the text box.