Specifications

Operation in Receiver Mode R&S FSH
1145.5973.12 2.20 E-10
Operation in Receiver Mode
(Available only if the option R&S FSH-K3 is installed.)
To provide a means of measuring levels at a specific frequency, the R&S FSH offers the receiver mode
as an option (option R&S FSH-K3). With this option, the R&S FSH functions like a receiver that
measures the level at a predefined frequency.
Switching on the receiver mode:
¾ Press the MEAS key.
¾ Press the MEASURE softkey.
The R&S FSH opens the menu for measurement
functions.
¾ Using the cursor keys or the rotary knob, select
RECEIVER and confirm with the ENTER key or
MEASURE softkey.
The R&S FSH activates the receiver mode and measures the level at the specified frequency.
The most important settings for the measurement
parameters are provided directly in the main menu of
the receiver mode, or they can be entered using the
corresponding keys.
Setting the frequency:
¾ Press the FREQ softkey in the main menu of the receiver mode.
¾ Using the rotary knob or the cursor keys, adjust the frequency, or, using the numeric keys, enter a
new frequency and confirm the entry with the ENTER key.
You can also enter the frequency by using the FREQ key.
Selecting the frequency step size:
The frequency resolution in the receiver mode is 100 Hz. The tuned step size can be changed as
required for the application