User manual

Waveform Setup
R&S
®
Scope Rider RTH
71User Manual 1326.1578.02 ─ 04
Range
Selects how the time limit of the window is defined. Time conditioning is available for
the vertical conditions "Stay Inside" and "Stay Outside".
"Longer"
Triggers if the signal crosses the upper or lower level after the speci-
fied "Width" time is reached.
"Shorter"
Triggers if the signal crosses the upper or lower level before the
specified "Width" time is reached.
"Equal"
Triggers if the signal stays inside or outside the vertical window limits
for the time "Width" "±Tolerance".
"Unequal"
Triggers if the signal stays inside or outside the vertical window limits
for a time unequal to "Width" "±Tolerance"
"Inside"
Triggers if the signal stays inside or outside the vertical window limits
at least for the time "Min Width" and for "Max Width" at the most.
"Outside"
"Outside" is the opposite definition of "Inside". The trigger occurs if
the signal stays inside or outside the vertical window limits for a time
shorter than "Min Width" or longer than "Max Width".
Remote command:
TRIGger:WINDow:TIME on page 184
TRIGger:WINDow:WIDTh on page 185
TRIGger:WINDow:DELTa on page 186
2.5.15 Protocol Trigger (R&S RTH-K1 and R&S RTH-K2)
The protocol trigger requires at least one of the serial protocol options. For protocol
setup and trigger settings, see
Trigger