User manual

Waveform Setup
R&S
®
Scope Rider RTH
64User Manual 1326.1578.02 ─ 04
±Tolerance
Sets a tolerance range Δt to the specified Time if the comparison range is equal or
unequal.
Remote command:
TRIGger:SLEW:DELTa on page 180
Min Time / Max Time
Set the lower and upper time limits if "Inside" or "Outside" is set for comparison.
2.5.10 Data2Clock Trigger (R&S RTH-K19)
With the Data2Clock trigger - also known as setup/hold trigger - you can analyze the
relative timing between two signals: a data signal and the synchronous clock signal.
Many systems require, that the data signal must be steady for some time before and
after the clock edge, for example, the data transmission on parallel interfaces.
The reference point for the time measurement is defined by clock level and clock edge.
SETUP key > "Trigger Type" = "Data2Clk"
Description of settings
Figure 2-15: Data2Clock Trigger
Clock Source
Selects the input channel of the clock signal.
Remote command:
TRIGger:DATatoclock:CSOurce[:VALue] on page 181
Trigger