Instructions

Table Of Contents
Fernsteuerbefehle
R&S
®
Scope Rider RTH
308Bedienhandbuch 1326.1578.03 ─ 10
Time scale Min./max. horizontal position
1 ns/div to 100 µs/div ±2 s
200 µs/div to 2 s/div Time scale * 20000
5 s/div – 500 s/div 100000 s
TIMebase:REFerence <ReferencePoint>
Defines the time reference point in the diagram.
Parameter:
<ReferencePoint> Position of the reference point in percent of the screen width.
Available values are:
10: on the left side of the screen
50: in the middle of the screen
90: on the right side
*RST: 50
Std-einheit: %
ACQuire:POINts:ARATe?
Returns the number of recorded analog waveform points per second.
Rückgabewerte:
<ADCSampleRate> Bereich: 1.25E+9 bis 5E+9
Inkrement: 1
*RST: 1.25E+9
Std-einheit: Sa/s
Verwendung: Nur Abfrage
15.3.4 Erfassungssteuerung
RUN............................................................................................................................. 308
STOP........................................................................................................................... 309
ACQuire:MODE............................................................................................................. 309
ACQuire:AVERage:COUNt............................................................................................. 309
ACQuire:ARESet:IMMediate........................................................................................... 309
ACQuire:POINts[:VALue]?.............................................................................................. 310
ACQuire:RESolution?.....................................................................................................310
RUN
Starts the continuous acquisition.
Verwendung:
Ereignis
Grundeinstellungen