Instructions

Table Of Contents
Fernsteuerbefehle
R&S
®
Scope Rider RTH
309Bedienhandbuch 1326.1578.03 ─ 10
STOP
Stops the running acquistion.
Verwendung: Ereignis
ACQuire:MODE <AcquisitionMode>
Defines how the waveform is built from the captured samples.
Parameter:
<AcquisitionMode> SAMPle | PDETect | HRESolution | AVERage | ENVelope
SAMPle
One of n samples in a sample interval is recorded as waveform
point, the other samples are discarded.
PDETect
The minimum and the maximum of n samples are recorded as
waveform points, the other samples are discarded.
HRESolution
The average of n captured sample points is recorded as one
waveform point.
AVERage
The average is calculated from the data of the current acquisi-
tion and a number of acquisitions before. The number of acquisi-
tions for average calculation is defined with ACQuire:
AVERage:COUNt.
ENVelope
The minimum and maximum values in an sample interval over a
number of acquisitions are saved. The most extreme values of
all acquisitions build the envelope.
*RST:
SAMPle
ACQuire:AVERage:COUNt <NoOfAvgs>
Sets the number of waveforms used to calculate the average waveform.
Parameter:
<NoOfAvgs> Bereich: 2 bis 8192
Inkrement:
2
N
(N = 1 .. 13)
*RST: 2
ACQuire:ARESet:IMMediate
Restarts the envelope and average calculation.
Verwendung:
Ereignis
Grundeinstellungen