Instructions

Table Of Contents
Fernsteuerbefehle
R&S
®
Scope Rider RTH
333Bedienhandbuch 1326.1578.03 ─ 10
TRIGger:INTerval:RANGe <Condition>
Defines how the time range of an interval is defined.
Parameter:
<Condition> LONGer | SHORter | EQUal | NEQual | WITHin | OUTSide
LONGer | SHORter
Triggers on intervals shorter or longer than an interval that is set
using TRIGger:INTerval:WIDTh.
EQUal | NEQual
Triggers on intervals equal or not equal a given interval width
that is set using TRIGger:INTerval:WIDTh. In addition, a
tolerance can be set around the specified width using
TRIGger:INTerval:DELTa.
WITHin | OUTSide
Triggers on intervals inside or outside a given range. The range
is set using ... and ....
*RST:
LONGer
TRIGger:INTerval:WIDTh <Duration>
Sets the time between two pulses for comparisons EQUal, UNEQual, SHORter, and
LONGer.
Parameter:
<Duration> Bereich: 800E-12 bis 10000
Inkrement: 100E-9
*RST: 5E-9
Std-einheit: s
TRIGger:INTerval:DELTa <Tolerance>
Sets a range Δt to the specified width, which is defined using TRIGger:INTerval:
WIDTh. The command is relevant for comparisons EQUal and UNEQual.
Parameter:
<Tolerance> Bereich: 0 bis 5000
Inkrement: 500E-12
*RST: 0
Std-einheit: s
TRIGger:INTerval:MINWidth <MinDuration>
Sets the lower time limit of the interval for comparisons WITHin and OUTSide.
Grundeinstellungen