391Bedienhandbuch 1326.1578.03 ─ 10
METer:SENSe:FUNCtion <MeasType>
Sets the measurement type and activates it. All measurement attributes of the previous
function (range, resolution, etc.) are remembered. If you return to the previous function,
the measurement attributes are restored.
Changing the measurement type disables scaling, limit testing, histogram, statistics,
and trend chart data collection: CALC:<function>:STAT is set OFF.
Einstellparameter:
<MeasType> 'CAPacitance | CONTinuity | CURRent:AC | CURRent[:DC] |
CURRent:AD | DIODe | FREQuency | RESistance |
TEMPerature | VOLTage:AC | VOLTage[:DC] | VOLTage:AD'
String parameter
*RST: VOLTage [:DC]
Beispiel:
MET:SENS:FUNC "VOLT:AC"
MET:CONF:VAL?
<--"VOLT:AC 1"
Verwendung: Nur Einstellung
METer:CONFigure:VALue?
Returns the actual measurement type (short form) and the range.
Rückgabewerte:
<Configuration> String parameter
Beispiel:
MET:CONF:VAL?
<--"VOLT:DC 100"
The DMM is set to DC voltage measurement and range 100 V.
Verwendung: Nur Abfrage
METer<m>:TRIGger:MODE <TriggerMode>
Defines how long the instrument measures. To start the measurement again, use RUN
or the appropriate METer<m>:MEASure:<function> command.
Suffix:
<m>
.
1..4
R&S RTH1002: always 1, omit the suffix
R&S RTH1004: Selects the voltmeter.
Parameter:
<TriggerMode> AUTO | SINGle
AUTO
The instrument performs continuous measurements.
SINGle
The instrument performs a single measurement.
*RST: AUTO
Digitalmultimeter (R&S RTH1002)