METer:MEASure:CAPacitance? [<ExpectedValue>]
Configures the capacitance measurement and returns the result.
Abfrageparameter:
[<ExpectedValue>] <numeric value> | MIN | MAX | DEF | 'AUTO'
Optional parameter, expected measurement result or range. The
instrument sets the appropriate measurement range.
See: METer:CONFigure:CAPacitance.
Rückgabewerte:
<Result> Numeric value
Verwendung: Nur Abfrage
METer:MEASure:CURRent:AC?
[<ExpectedValue>]
METer:MEASure:CURRent:DC? [<ExpectedValue>]
Configures the current measurement and returns the result.
Abfrageparameter:
[<ExpectedValue>] <numeric value> | MIN | MAX | DEF | 'AUTO'
Optional parameter, expected measurement result or range. The
instrument sets the appropriate measurement range.
See: METer:CONFigure:CURRent:DC.
Rückgabewerte:
<Result> Numeric value
Beispiel:
:METer:MEASure:CURRent:DC? 15
<-- 13.4907681509
Verwendung: Nur Abfrage
METer:MEASure:RESistance? [<ExpectedValue>]
Configures the resistance measurement and returns the result.
Abfrageparameter:
[<ExpectedValue>] <numeric value> | MIN | MAX | DEF | 'AUTO'
Optional parameter, expected measurement result or range. The
instrument sets the appropriate measurement range.
See: METer:CONFigure:RESistance.
Rückgabewerte:
<Result> Numeric value
Verwendung: Nur Abfrage
METer:MEASure:VOLTage:AC? [<ExpectedValue>]
METer:MEASure:VOLTage:DC? [<ExpectedValue>]
Configures the voltage measurement and returns the result.
Digitalmultimeter (R&S
RTH1002)