User manual

Remote Control Commands
R&S
®
Scope Rider RTH
163User Manual 1326.1578.02 ─ 04
CHANnel<m>:OFFSet <Offset>
Sets an offset voltage that is added to correct an offset-affected signal. The value is
included in measurement results. The signal is shifted in relation to the ground level by
the offset value. Negative offset values move the waveform down, positive values
move it up.
Suffix:
<m>
.
1|2 (RTH1002) , 1..4 (RTH1004)
Parameters:
<Offset> Range: -400 to 400
Increment: 0.5
*RST: 0
Default unit: V
CHANnel<m>:COUPling <Coupling>
Selects the connection of the indicated channel.
Suffix:
<m>
.
1|2 (RTH1002) , 1..4 (RTH1004)
Parameters:
<Coupling> DCLimit | ACLimit
DCLimit
The signal passes the input unchanged.
ACLimit
A high-pass filter removes the DC offset voltage from the input
signal if the DC component of a signal is of no interest.
*RST: DCLimit
CHANnel<m>:BANDwidth <BandwidthLimit>
Selects the bandwidth limit for the indicated channel.
Suffix:
<m>
.
1|2 (RTH1002) , 1..4 (RTH1004)
Waveform Setup