TRIGger:WIDTh:MIN <MinDuration>
Sets the lower limit for the pulse width if TRIGger:WIDTh:RANGe is set to WITHin or
OUTSide.
Parameters:
<MinDuration> Range: 800E-12 to 10000
Increment: 100E-9
*RST: 5E-9
Default unit: s
12.3.5.5 Pattern Trigger
See also Chapter 2.5.6, "Pattern Trigger (R&S RTH-K19)", on page 57.
TRIGger:PATTern:STATe[:CHANnel<1..22>]....................................................................174
TRIGger:PATTern:STATe:COMBination...........................................................................174
TRIGger:PATTern:WIDTh:RANGe...................................................................................175
TRIGger:PATTern:TIMeout[:TIME]...................................................................................175
TRIGger:PATTern:WIDTh[:WIDTh].................................................................................. 175
TRIGger:PATTern:WIDTh:DELTa.................................................................................... 176
TRIGger:PATTern:STATe[:CHANnel<1..22>] <State>
Sets the state of each input channel. The channel is specified by the channel suffix:
1..4: analog channel 1 to 4
8..15: digital channels D0 to D7
5..7 and 16..22: not available
The logical combination of the channel states is defined by TRIGger:PATTern:
STATe:COMBination.
Parameters:
<State> ONE | ZERO | DONTcare
ONE
The signal value is above the defined threshold.
ZERO
The signal value is below the defined threshold.
DONTcare
The signal state does not matter.
*RST: DONTcare
TRIGger:PATTern:STATe:COMBination <ChCombination>
Sets the logical combination for all active channels. The required state of each channel
is defined by TRIGger:STATe:CHANnel<1..22>.