User manual

Remote Control Commands
R&S
®
Scope Rider RTH
176User Manual 1326.1578.02 ─ 04
See TRIGger:PATTern:WIDTh:RANGe on page 175.
Parameters:
<Duration> Range: 800E-12 to 10000
Increment: 100E-9
*RST: 5E-9
Default unit: s
TRIGger:PATTern:WIDTh:DELTa <Tolerance>
Sets a range Δt to the pattern width that is defined by TRIGger:PATTern:WIDTh[:
WIDTh].
Parameters:
<Tolerance> Range: 0 to 5000
Increment: 500E-12
*RST: 0
Default unit: s
12.3.5.6 State Trigger
See also Chapter 2.5.7, "State Trigger (R&S RTH-K19)", on page 59.
TRIGger:STATe:CHANnel<1..22>................................................................................... 176
TRIGger:STATe:COMBination.........................................................................................177
TRIGger:STATe:CSOurce[:VALue].................................................................................. 177
TRIGger:STATe:CSOurce:EDGE.....................................................................................177
TRIGger:STATe:CHANnel<1..22> <State>
Sets the state of each input channel. The channel is specified by the channel suffix:
1..4: analog channel 1 to 4
8..15: digital channels D0 to D7
5..7 and 16..22: not available
The logical combination of the channel states is defined by TRIGger:STATe:
COMBination.
Parameters:
<State> ONE | ZERO | DONTcare
ONE
The signal value is above the defined threshold.
ZERO
The signal value is below the defined threshold.
DONTcare
The signal state does not matter.
*RST: DONTcare
Waveform Setup