User manual

Waveform Setup
R&S
®
Scope Rider RTH
58User Manual 1326.1578.02 ─ 04
Figure 2-11: Pattern definition
The current threshold is displayed for each channel. Tap "Find Level" to determine a
useful threshold for all active channels.
The switches define the required state of each channel and set the logical combination:
"1"
The signal value is above the defined threshold.
"0"
The signal value is below the defined threshold.
"X"
The signal state does not matter.
"AND"
If all defined states are true, the logical result of the pattern definition
is 1 (true).
"OR"
If at least one of the defined states is true, the logical result of the pat-
tern definition is 1 (true).
Remote command:
TRIGger:PATTern:STATe[:CHANnel<1..22>] on page 174
TRIGger:PATTern:STATe:COMBination on page 174
Range
Adds additional time limitation to the defined pattern.
"None"
No time limit is set. If the defined pattern is true, the pattern trigger
occurs.
"Timeout"
Defines a minimum time during which the signals match the pattern
definition.
"Longer"
If the pattern is true longer than the "Pattern Width" time, the trigger
occurs.
"Shorter"
If the pattern is true for a time shorter than "Pattern Width", the trigger
occurs.
Trigger