User manual

Waveform Setup
R&S
®
Scope Rider RTH
62User Manual 1326.1578.02 ─ 04
Remote command:
TRIGger:RUNT:WIDTh on page 178
±Tolerance
Sets a tolerance range Δt to the specified Runt Width if the comparison range is equal
or unequal.
Remote command:
TRIGger:RUNT:DELTa on page 178
Min Runt Width / Max Runt Width
Set the lower and upper time limits if "Inside" or "Outside" is set for comparison.
2.5.9 Slew Rate Trigger (R&S RTH-K19)
The slew rate trigger, also known as transition trigger, triggers on edges if the transition
time from the lower to higher voltage level (or vice versa) is shorter or longer as
defined, or outside or inside a specified time range.
The slew rate trigger finds slew rates faster than expected or permissible to avoid over-
shooting and other interfering effects. It also detects very slow edges violating the tim-
ing in pulse series.
UL
L L
min
max
Figure 2-14: Slew rate trigger, transition time inside a range (t > min time AND t < max time)
SETUP key > "Trigger Type" = "Slew rate"
Trigger