User manual

Waveform Setup
R&S
®
Scope Rider RTH
67User Manual 1326.1578.02 ─ 04
Remote command:
TRIGger:SPATtern:PATTern on page 183
Bit Order
Defines if the data words start with msb (most significant bit) or lsb (least significant
bit).
Remote command:
TRIGger:SPATtern:ORDer on page 183
2.5.12 Timeout Trigger (R&S RTH-K19)
The timeout trigger checks if the signal stays above or below the trigger level for a
specified time lapse. In other words, the trigger occurs if the signal does not cross the
trigger level during the specified time.
SETUP key > "Trigger Type" = "Timeout"
Trigger