User manual

Waveform Setup
R&S
®
Scope Rider RTH
69User Manual 1326.1578.02 ─ 04
Description of settings
Figure 2-18: Interval Trigger
Slope
Sets the edge for the trigger. You can analyze the interval between positive edges or
between negative edges.
Remote command:
TRIGger:INTerval:SLOPe on page 186
Range
Defines how the time range of an interval is defined. The comparisons are the same as
for the width trigger, see "Range" on page 52.
Remote command:
TRIGger:INTerval:RANGe on page 186
Interval Width
Sets the time between two pulses for comparisons equal, unequal, shorter, and longer.
Remote command:
TRIGger:INTerval:WIDTh on page 187
±Tolerance
Sets a tolerance range Δt to the specified Interval Width if the comparison range is
equal or unequal.
Remote command:
TRIGger:INTerval:DELTa on page 187
Min Interval Width / Max Interval Width
Set the lower and upper time limits if "Inside" or "Outside" is set for comparison.
Trigger