User manual

Waveform Analysis
R&S
®
Scope Rider RTH
94User Manual 1326.1578.02 ─ 04
b) Set the index of the first and the last segment that you want to see.
See "Start Acquisition / Stop Acquisition" on page 92.
c) In the history player window, tap the Play button.
4. To play back the segments repeatedly.
a) Enable the Repeat button.
b) Tap the Play button.
5. To access a particular segment, you can:
Drag the slider in the history player window, until the required index number is
shown.
Use the Step Forward and Step Back buttons to show the next segment.
Remote command:
CHANnel<m>:HISTory:PLAY on page 199
CHANnel<m>:HISTory:REPLay on page 199
CHANnel<m>:HISTory:CURRent on page 199
3.7.3 Analyzing History Data
You can analyze the stored segments using the following possibilities:
Switch analog channels on and off if they were active during acquisition.
Change vertical scale and position of captured channels.
Use the zoom.
Save screenshots.
Perform cursor measurements.
Perform automatic measurements.
Use the mathematical functions.
Switch logic channels on and off if at least one channel was active during acquisi-
tion.
Enable and disable the serial bus if it was active during acquisition.
History (Option R&S RTH-K15)