Mask Testing
R&S
®
Scope Rider RTH
97User Manual 1326.1578.02 ─ 04
2. If you want to test a mathematical (math) waveform, generate a math waveform,
see Chapter 3.4, "Mathematics", on page 84.
3. Select the "Mask" menu.
4. Select the mask channel that is assigned to the tested waveform.
5. To activate the mask, tap "State".
Note: If the selected mask was used before, the stored mask is displayed.
6. To set the mask width, enter "Width X" and "Width Y".
7. Select the "Action on Violation".
This action is applied to all masks.
8. To create the mask, tap "Create Mask".
9. To start and stop mask testing, press the RUN STOP key.
Analyzing the mask test
To view the details of the waveform at a mask test violation, you can use the zoom
function, see Chapter 3.1, "Zoom", on page 72. Zooming is only possible while the
mask test is stopped.
4.4 Mask Settings
Access: "Mask" menu
Mask Settings