User manual

Multimeter Measurements
R&S
®
Scope Rider RTH
107User Manual 1326.1578.02 ─ 04
7 = Warning sign, shows up if the measured value is higher than 30 V
8 = Bar graph displaying the measured value
9 = Measurement type
10 = Activates or deactivates relative measurement
11 = Restarts the measurement and resets all values
Remote command:
METer<m>:READ? on page 205
METer<m>:FETCh? on page 206
In meter mode, the keys behave slightly different from scope mode:
The vertical RANGE and POS keys adjust the measurement range.
The MEAS key opens the "Meter" menu.
The following keys work as usual: FILE,
, , PRESET, AUTOSET, MODE,
BACK.
The CH keys switch back to the scope mode.
All other keys do not work.
5.2.3 Running Multimeter Measurements
1. Connect the test leads to the meter inputs.
2. Select the "Meter" mode.
3. Select the measurement type (no. 9 in Figure 5-5).
4. To adjust the measurement range, press the RANGE keys.
5. For current and temperature measurements, adjust the "Probe Setting".
6. For continuity measurement, adjust the "Resistance Threshold".
7. If you want to get relative result values:
a) Activate "Relative" (no. 10).
b) To set a user-defined reference value, open the "Meter" menu, activate "Refer-
ence", and enter the value in "Reference Manual".
8. To stop the meter measurement, press the RUN STOP key.
5.2.4 Settings
Access: "Meter" menu
Digital Multimeter (R&S RTH1002)